We bring up the ChipScope software
and set up the trigger condition as
"macro" with EXT as the source. In
addition, the capture mode is set up
as "One Shot" (default). We enable
the “Ext. Out” signal for synchroniza-
tion purposes (we are probing this
signal with both analyzers). The ILA
is now set up to trigger when its
external trigger input goes high.
Starting the ILA at this point will
arm it and force it to wait for the
16700 Series system to trigger.
Figure 8 shows the trigger position at
128 (out of 256 samples). The user
can set the trigger to any valid loca-
tion. The example screen also drives
the external trigger-out signal even
though the unit is not triggering any
other analyzers. This signal will be
used to correlate the data sets from
the Xilinx ILA and the Agilent logic
analysis module.
Figure 8. Xilinx ChipScope armed and ready for external trigger
Take the Measurement
The Xilinx ILA can now be started
from the ChipScope interface. It will
arm and wait for the trigger from the
Agilent analysis module. The Agilent
16700 Series can then be started.
Once its trigger condition is detected,
both the Xilinx ILA and Agilent
16700 Series system will finish and
display data. Figures 9 and 10 show
both analyzers having completed an
acquisition.
Figure 9. Agilent 16700 Series logic analysis system complete measurement