Agilent 75000 Series B
Agilent E1326B
5 1/2-Digit Multimeter
Service Manual
Enclosed is the Service Manual for the Agilent E1326B
5 1/2-Digit Multimeter. Insert this manual, along with
any other VXIbus manuals that you have, into the binder
that came with your Agilent Technologies mainframe.
*E1326-90017*
Manual Part Number: E1326-90017
Printed in Malaysia E0306
Vista de pagina 0
1 2 3 4 5 6 ... 78 79

Indice de contenidos

Pagina 1 - *E1326-90017*

Agilent 75000 Series BAgilent E1326B5 1/2-Digit MultimeterService Manual Enclosed is the Service Manual for the Agilent E1326B

Pagina 2

1General InformationIntroduction This Agilent E1326B Service Manual contains information required totest, adjust, troubleshoot, and repair the Agilent

Pagina 3

Warnings andCautionsThis section contains WARNINGS which must be followed for yourprotection and CAUTIONS which must be followed to avoid damage tothe

Pagina 4 - Safety Symbols

WARNINGCAPACITOR VOLTAGES. Capacitors inside the mainframe mayremain charged even when the mainframe has beendisconnected from its source of supply.U

Pagina 5 - Jim White, QA Manager

MultimeterSpecifications Multimeter specifications are listed in Appendix A of the AgilentE1326B/E1411B User’s Manual. These specifications are thepe

Pagina 6 - Manual Descriptions

Recommended Test EquipmentTable 1-1 lists the test equipment recommended for testing, adjustingand servicing the multimeter. Essential requirements fo

Pagina 7 - What’s in this Manual

NOTES:1-6 General Information

Pagina 8 - Contents

2InstallationIntroduction This chapter provides information to install the Agilent E1326Bmultimeter, including initial inspection, preparation for use

Pagina 9 - Chapter 8 - Service

Shipping the MultimeterIf you need to return the Agilent E1326B multimeter to AgilentTechnologies, first remove any adapters or connectors before pack

Pagina 10 - General Information

3 Operating InstructionsIntroduction This chapter lists operating information for the Agilent E1326Bmultimeter, including:• Multimeter operation• Oper

Pagina 11 - Cautions

3. A "0" returned means no self-test failure, while "1", "2", "3", or "4"returned means a failure w

Pagina 13 - 1-4 General Information

4Verification TestsIntroduction The three levels of test procedures described in this chapter are used toverify that the Agilent E1326B multimeter:• i

Pagina 14 - Test Equipment

NOTEThe upper and lower limits in the Performance Test Record assume thetest equipment used is calibrated and operating at peak performance. Ifthis i

Pagina 15 - 1-6 General Information

3. Execute the multimeter functional verification test using the *TST?command. See the following example which uses an HP 9000 Series 300computer wit

Pagina 16 - Installation

Test 4-1: DC Voltage Test (Zero Volt Input)This test verifies DC Voltage accuracy on all five ranges with a zero volt input.Equipment Setup 1. Connec

Pagina 17 - Shipping

3. Repeat steps 1 and 2 for the following ranges: E1326BRange10%OverrangeInput0.91 V7.27 V58.1 V300 V1 V 8 V64 V N/A0 V0 V0 V0 V4. Remove power and

Pagina 18 - Operating Instructions

Test 4-2: DC Voltage Test (DCV Input)This test verifies DC Voltage accuracy on all five ranges with DCvoltage inputs.Equipment Setup 1. Connect the e

Pagina 19 - Self-Test

Test Procedure 1. Set the DC Standard (Datron 4708, Option 10) Output to 0.1 DCV.2. Set the Agilent E1326B range to 0.113 V (0.125 V wi

Pagina 20 - Verification Tests

100 PRINT "Set DC Standard to ";Volts(I);" VDC" 110 PRINT "Press Continue when ready" 120 PAUSE 130 CLE

Pagina 21 - Verification

WARNINGThe AC Standard (Datron 4708, Option 20) can producedangerous voltages which are present on the terminals. Do nottouch the front (or rear) pan

Pagina 22 - Example: Self-Test

Example: AC Voltage TestThis example performs an ACV test for a power line referencefrequency of 60 Hz. Change line 80 to OUTPUT 70903;"CAL:LFR 5

Pagina 23 - 4-4 Verification Tests

CertificationAgilent Technologies certifies that this product met its published specifications at the time of shipment from the factory. AgilentTechn

Pagina 24

Test 4-4: Resistance Test (4-Wire Ohms)This test verifies the 4-wire resistance accuracy of the 2kΩ, 131kΩ, and1MΩ ranges.NOTEThe DC Voltage performan

Pagina 25 - 4-6 Verification Tests

Test Procedure 1. Set the Resistance Standard (Datron 4708, Option 30) to 1 kΩ setting.2. Set the Agilent E1326B range to 1861Ω (2048Ω

Pagina 26

1 ! 4-Wire Ohms Performance Verification2 ! 10 OPTION BASE 1 20 DIM Range(3),Source(3),Read_meas(3),Limit(3),Ohms(3) 30 DATA 1861, 11

Pagina 27 - Test 4-3: AC Voltage Test

Performance Test RecordTable 4-1, Performance Test Record for the Agilent E1326B Multimeter,is a form you can copy and use to record performance veri

Pagina 28

Table 4-1. Performance Test Record for the Agilent E1326B Multimeter (Page 1 of 3)Test Facility:Name _____________________________________Address __

Pagina 29 - 4-10 Verification Tests

Table 4-1. Performance Test Record for the Agilent E1326B Multimeter (Page 2 of 3)Model _________________________________ Report No. ____________

Pagina 30 - Verification Tests 4-11

Table 4-1. Performance Test Record for the Agilent E1326B Multimeter (Page 3 of 3)Model ________________________________ Report No. _______________

Pagina 31 - 4-12 Verification Tests

NOTES:4-18 Verification Tests

Pagina 32 - Verification Tests 4-13

5AdjustmentsIntroduction This chapter contains procedures to adjust the Agilent E1326Bmultimeter for peak performance. For best performance, theinst

Pagina 33 - Test Record

DC VoltageAdjustments This procedure adjusts Agilent E1326B DC voltage measurementaccuracy.Equipment Setup1. Connect the equipment as shown in Figure

Pagina 34 - Verification Tests 4-15

Frame or chassis ground terminal—typicallyconnects to the equipment’s metal frame.Alternating current (AC).Direct current (DC). Indicates hazardous vo

Pagina 35 - 4-16 Verification Tests

NOTE*RST sets Auto Zero to ON and Power Line Cycles to 1.AdjustmentProcedure1. Set the DC Standard output to 7.7 V.2. Set the Agilent E1326B range t

Pagina 36 - 1-Year Specifications

displays an "adjustment complete" message. If a calibration error occurs,the program displays the calibration error and prompts you to repea

Pagina 37 - 4-18 Verification Tests

310 PRINT "Press Continue for the next adjustment" 320 PAUSE 330 CLEAR SCREEN 340 NEXT I 350 END AC Voltage AdjustmentsThis

Pagina 38 - Adjustments

• Power Line Cycles (PLC) ... 1• Line Freq Reference (CAL:LFR) ... 50 Hz or 60 HzNOTE*RST sets

Pagina 39 - DC Voltage

30 Re_try: !40 PRINT "Set AC Standard to 5.6 Vac at 1.0 kHz" 50 PRINT "Press Continue when ready" 60 PAUSE 70 C

Pagina 40 - Example: DC

NOTEThe DC Voltage adjustment and the AC Voltage adjustment MUST beperformed before the Resistance adjustment.Equipment Setup 1. Connect the equipme

Pagina 41 - 5-4 Adjustments

4. Repeat steps 1 through 3 using the following Agilent E1326B (and FRES:RANG) ranges, and Resistance Standard (and CAL:VAL) values. Agilent

Pagina 42 - AC Voltage

250 CLEAR SCREEN 260 GOTO Re_try 270 ELSE 280 PRINT "Adjustment complete with";Source(I);"Ohms source" 290

Pagina 43 - Example: AC

14 PacerOverrunErrorThe pacer is in use and the pacer rate is faster than themaximum A/D conversion rate based on integration time,autorange setting,

Pagina 44 - Resistance

5-12 Adjustments

Pagina 45

Declaration of Conformityaccording to ISO/IEC Guide 22 and CEN/CE NELEC EN 45014Manufacturer’s Name: Agilent Technologies, Inc.Loveland Manufacturing

Pagina 46

6Replaceable PartsIntroduction This chapter contains information to order replaceable parts for theAgilent E1326B multimeter. ExchangeAssembliesTable

Pagina 47 - Calibration

Reference*DesignatorAgilent PartNumberQty Description Mfr**CodeMfr Part NumberA2A2J101A2J102A2J103A2K104-K105A2K106A2XU104E1326-6650244702-616031252-3

Pagina 48 - Adjustments 5-11

Figure 6-1. Agilent E1326B Replaceable PartsReplaceable Parts 6-3

Pagina 49 - 5-12 Adjustments

Table 6-3. Agilent E1326B Code List of ManufacturersMfrCodeManufacturer Name Address00000007790677618873249312726428480717077591581

Pagina 50 - Replaceable Parts

7Manual ChangesIntroduction This chapter contains information to adapt this manual to instrumentsfor which the content does not directly apply. Since

Pagina 52 - Replaceable Parts 6-3

8ServiceIntroduction This chapter contains information to service the Agilent E1326Bmultimeter, including:• recommended repair strategy• troubleshooti

Pagina 53 - 6-4 Replaceable Parts

Service notes and other service literature for the Agilent E1326Bmultimeter may be available through Agilent. For information, contactyour nearest Ag

Pagina 54 - Manual Changes

Table 8-2. Self-Test ErrorsError Description Typical Causes1234Multimeter does not respond toself-testInvalid communication

Pagina 55 - 7-2 Manual Changes

Table 8-3. Agilent E1326B Tests/ChecksTest/Check ReferenceDesignatorCheck:Heat Damage - - - - - - - - - - Discolored PC boardsDa

Pagina 56 - Required

Agilent 75000 Series B Service DocumentationSuggested Sequence to Use ManualsManual DescriptionsInstallation and Getting Started Guide. Cont

Pagina 57 - Techniques

Testing A1/A2 PCAsTo test the A1 Outguard PCA and the A2 Inguard PCAs, removemainframe power and remove the multimeter from the mainframe.Then, remove

Pagina 58 - Assemblies

Repair/MaintenanceGuidelinesThis section gives guidelines to repair and maintain the Agilent E1326Bmultimeter, including:• ESD precautions• Removing A

Pagina 59 - 8-4 Service

components.• After you remove an assembly from the multimeter, place theassembly on a conductive surface to guard against ESD damage.Do not stack asse

Pagina 60 - Service 8-5

Removing A2Inguard PCAUse the following steps to remove the A2 A/D Inguard printed circuitassembly (PCA) from the A1 Outguard PCA. See Figure 8-1 for

Pagina 61 - Guidelines

Soldering Printed CircuitBoardsThe etched circuit boards in the multimeter have plated-through holesthat allow a solder path to both sides of the insu

Pagina 62 - Service 8-7

NOTES:8-10 Service

Pagina 63 - Binding Posts

ACalculating Multimeter AccuracyIntroduction This appendix shows how multimeter accuracy, measurementuncertainty, and test accuracy ratio (TAR) values

Pagina 64 - Safety Checks

Test Accuracy Ratio (TAR) DefinitionTest Accuracy Ratio (TAR) is the ratio of multimeter accuracy tomeasurement uncertainty. For the Agilent E1326B m

Pagina 65 - 8-10 Service

input the Low Limit in Table 4-1 = 6.9986 Volts and the High Limit =7.0014 Volts. AC VoltageAccuracy EquationsFrom Appendix A of the Agilent E1326B/E

Pagina 66

Measurement UncertaintyCalculationsMeasurement uncertainties for the Datron 4708 source are calculatedusing the 90-day accuracy specifications in the

Pagina 67 - Calculations

What’s in this ManualManual Overview This manual shows how to service the Agilent E1326B 5 1/2-Digit Multimeter. Additional manuals whichmay be requ

Pagina 68

Calculate ACVMeasurement Uncertainty From Section 6 - Specifications of the Datron 4708 User’s Handbook, ACVoltage (Option 20) Accuracy (90 days since

Pagina 69 - Uncertainty

Test AccuracyRatio (TAR)CalculationsFor the Agilent E1326B multimeter Test Accuracy Ratio (TAR) =[High Limit - Input Value]/Measurement Uncertainty w

Pagina 70

BVerification Tests - C ProgramsFunctionalVerificationTestThis program is designed to do the Functional Verification Test found inChapter 4 - Verifica

Pagina 71 - Ratio (TAR)

PerformanceVerificationTestsThese programs are designed to do the Performance Verification Testsfound in Chapter 4 - Verification Tests.Example: Zero

Pagina 72

Example: DCVoltage TestThis test performs a DC Voltage test for positive input DC volts and apower line reference frequency of 60 Hz. /* DC Voltage T

Pagina 73

Example: ACVoltage TestThis example performs an AC voltage test for a power line referencefrequency of 60 Hz./* AC Voltage Test E1326B */

Pagina 74 - Voltage Test

Example:Resistance TestThis example performs a 4-wire ohms resistance test. The program alsocalculates the Upper and Lower Limit values for the ACTUA

Pagina 75

Adjustments These programs are designed to do the adjustments found in Chapter 5 -Adjustments.DC VoltageAdjustmentsThis example performs DC Voltage ad

Pagina 76 - Resistance Test

AC VoltageAdjustmentsThis example performs an AC Voltage adjustment for a power linereference frequency of 60 Hz and an input of 5.6 Vac at 1 kHz. If

Pagina 77

ResistanceAdjustmentsThis example performs a 4-wire ohms resistance adjustment for a powerline reference frequency of 60 Hz. If a calibration error o

Pagina 78

ContentsChapter 1 - General InformationIntroduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Pagina 79

Chapter 6 - Replaceable PartsIntroduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .