Vista de pagina 182
2-58
Assembly Replacement
AT1 115 dB Attenuator (Option 1EM)
AT1 115 dB Attenuator (Option 1EM)
Tools Required
T-10 driver
5/16” open ended wrench
Removal Procedure
Refer to Figure 2-25 for this procedure.
1. Disconnect the power cord.
2. Remove the covers from the signal generator. Refer to "Outer Instrument Cover" on page 2-5
and "Inner Instrument Cover" on page 2-7.
3. Position the signal generator with the A31 Motherboard facing you so that the RF deck is on top.
4. Disconnect the AT1W1 ribbon cable from A26J13.
5. Using the 5/16” open ended wrench, disconnect the W72 and W73 RF cables from the
AT1 115 dB Attenuator.
6. Using the T-10 driver, remove the two screws that attach the AT1 115 dB Attenuator to the side
panel.
7. Remove the AT1 155 dB Attenuator.
8. Using the T-10 driver, remove the sheet metal from the attenuator.
Replacement Procedure
Reverse the order of the removal procedure.
Torque all T-10 screws to 9 in-lbs.
Torque all RF connectors to 8 in-lbs.
Perform the post-repair adjustments and performance tests that pertain to this removal
procedure.