Agilent Technologies EXG Manual de usuario Pagina 70

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 98
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 69
Identifying Transmission Problems
7-8
Transmission stopped
unexpectedly. X of N ECGs sent.
Modem was disconnected.
Press any key to continue.
Problem with telephone line Check that the modem is
connected to the telephone line.
Verify that the telephone line is
working.
Transmission stopped
unexpectedly. X of N ECGs sent.
Remote site stopped
communication (nn). Press any
key to continue.
Communication speed of the
remote device does not match
that of the cardiograph, or
remote site modem malfunction.
Call the remote location to
verify that the communication
speed is correct and that the
modem is functioning correctly.
Reduce the communication
speed of both the cardiograph
and remote site modems.
Table 7-5 Transmission Problems and Solutions(Continued)
Message Likely Cause Possible Solutions