Operation
26
PRBS Generator 10 Gb/s User Guide
Figure 7. Verification setup
Correct pattern length is determined by confirming that the tone spacing on
the spectrum analyzer equals (Output_Bit_Rate/(2
n
-1)), where n is the
specific pattern length. As an example, at 10 Gb/s the tone spacing for a
(2
31
-1) pattern should be ~ 4.66 Hz (refer to Table 4 for tone spacings).