Instrument Location and Rack Mounting Requirements
Instrument Location and Rack Mounting Requirements
Locating the instrument
Make sure that the fan inlet and exhaust vent areas on the sides of the instrument are
not obstructed. The minimal required clearance is 2 inches. Airflow restrictions
cause additional airflow noise and cause the fans to speed up so they can draw in
enough air for the required cooling. This results in excessive audible noise.
Cooling and rack mounting
Do not rack mount the receiver side-by-side with any other instrument with
side-by-side ventilation. Make sure the exhaust air from the first instrument is
directed away from the inlet of the second unit. If the pre-heated air from the first
instrument is directed into the second instrument, it can cause excessive operating
temperatures in the second unit and can cause instrument failures. The instrument
draws air in from the left side and exhausts air from the right side.
CAUTION When installing the product in a cabinet, the convection into and out of the product
must not be restricted. The ambient temperature (outside the cabinet) must be less
than the maximum operating temperature of the product by 4° C for every 100 watts
dissipated in the cabinet. If the total power dissipated in the cabinet is greater than
800 watts, then forced convection must be used.