USER’S GUIDE AGILENT 6010A, 6011A, 6012B, 6015A, 6023A and 6028A DC AUTORANGING POWER SUPPLY Agilent Part Num
10 General InformationOptionDescription100 Input power: 100 Vac + 6%, -10%; 48-63 Hz single phase.120 Input power: 120 Vac +6%, -13%. 48-63 Hz sing
General Information 11PSI boards to lie on table outside unit, and control board test connector.5060-0138 GP-IB connector non-metric to metric conv
12 General InformationTable 1-1 Performance SpecificationsAgilent Technologies Model 6010A 6011ADC Output: Voltage, current and power spans indicat
General Information 136012B 6015A 6023A 6028A NOTES.0-60 V 0-500 V 0-20 V 0-60 V0-50 A 0-5 A 0-30 A 0-10 A1000-1200 W 1000-1050 W 200-242 W 200-242
14 General InformationTable 1-2 Supplemental Characteristics (continued)DC Floating Voltage: Either output terminal may be floated up to the follow
General Information 15Model Agilent6010AAgilent6011AAgilent6012BAgilent6015AAgilent6023AAgilent6028AVp1 200 V 20 V 60 V 500 V 20 V 60 VIp1 5 A 50 A
16 General InformationFigure 1-2. Output Impedance
Installation 172InstallationIntroductionThis section contains instructions for checking and repackaging the supply; bench or rack mounting, connect
18 InstallationOutline DiagramsFigure 2-1 illustrates the outline shape and dimensions of the cabinet.6023A ONLY6028AONLYFigure 2-1. Outline Diagram
Installation 19Power ConnectionCaution: Connection of this supply to an ac power source should only be performed by an electrician or otherqualifie
2 CERTIFICATION Agilent Technologies certifies that this product met its published specifications at time of shipment from the factory. Agilent Techn
20 InstallationThe offset pin on the standard power cable three-prong connector is the ground connection. If a two-contact receptacle isencountered,
Installation 21Figure 2-2. Power-Cord Plug ConfigurationsAgilent Models 6023A and 6028A.Line voltage conversion is accomplished via three component
22 InstallationAgilent 6010A, 6011A, 6012B, 6015AAgilent 6023A, 6028AFigure 2-3. Line Voltage Conversion Components
Installation 23AC Line Impedance CheckThe power supply is designed for proper operation with line impedances typically found in ac power lines. Ho
24 InstallationRear Panel Screw Sizes and Part NumbersRefer to the following list if you need to replace any of the rear panel connection hardware:A
Operating Instructions 253Operating InstructionsIntroductionThis section describes the operating controls and indicators, turn-on checkout procedure
26 Operating InstructionsTable 3-1. Controls and IndicatorsNumber Controls/Indicators Description Page1 LINE Switch Pressing at the top of the switch
Operating Instructions 27a. Check that the rear-panel mode switches are set as shown in Figure 3-3.b. Check that + lead is connected to +S and the
28 Operating InstructionsInitial Setup and InterconnectionsWARNING: Turn off input ac power before changing any rear-panel connection and make certai
Operating Instructions 29Table 3-2. Maximum Wire Lengths To Limit Voltage DropsWire Size Resistivity Maximum Length In Meters (Feet)ToLimit Voltage
3Safety SummaryThe following general safety precautions must be observed during all phases of operation of this instrument. Failure tocomply with thes
30 Operating InstructionsEither positive or negative voltages can be obtained from the supply by grounding one of the output terminals. It is best t
Operating Instructions 31circuits within the supply the power supply output drops to zero and the UNREGULATED indicator turns on. The powersupply c
32 Operating Instructionscrosses the operating locus at point 1. Point 1 is on the part of the operating locus defined by the voltage setting, so th
Operating Instructions 33Figure 3-5. Overrange OperationConstant Voltage OperationBy pressing the DISPLAY SETTINGS pushbutton switch you can observe
34 Operating InstructionsConstant Current OperationTo set up the power supply for constant current operation:a. With power supply turned off, connec
Operating Instructions 35Figure 3-6. Remote Voltage SensingBecause the sensing leads carry only a few milliamperes, the wires used for sensing can b
36 Operating InstructionsAnalog ProgrammingThese instruments can obtain their output voltage and current programming information from three distinct
Operating Instructions 37Note: If the programming terminals (VP to P) become open circuited during resistance programming, theoutput voltage will r
38 Operating InstructionsFigure 3-9. Optional Voltage Divider for Program SourceConstant Current Output, Resistance Control.The setup shown in Figure
Operating Instructions 39Constant Current Output, Voltage Control.The setup shown in Figure 3-11 allows the output current to be varied by using an
4SAFETY SYMBOLSDirect currentAlternating currentBoth direct and alternating currentThree-phase alternating currentEarth (ground) terminalProtective ea
40 Operating InstructionsFigure 3-12. Auto-Parallel OperationSetting Voltage and Current. Program the slave unit's output voltage above the maste
Operating Instructions 41Caution: When two supplies are operated in series, they should be programmed to the same voltage to preventpossible damage
Appendix A 43A100 VAC Input Power Option 100General InformationDescriptionOption 100 is a modification of the power supply that involves changing th
44 Appendix ASection 3 Manual ChangesOn page 30 under “Overvoltage Protection”, change 107% to 90%.On page 36 under “Analog Programming”, in the 3rd
Index 45IndexAairflow...
Index 46Rremote sensing ...
Agilent Sales and Support Office 47Agilent Sales and Support OfficeFor more information about Agilent Technologies test and measurement products, appl
48 Manual Updates The following updates have been made to this manual since the print revision indicated on the title page. 2/01/00 All references
5 DECLARATION OF CONFORMITY According to ISO/IEC Guide 22 and CEN/CENELEC EN 45014 Revision: B.00.00 Issue Date: Created on 11/24/2003 3:35 PM Doc
6Acoustic Noise StatementHerstellerbescheinigungDiese Information steht im Zusammenhang mit den Anforderungen derMaschinenlärminformationsverordnung v
Table Of Contents1. General InformationIntroduction ...
Multiple-Supply Operation.........3
General Information 91General InformationIntroductionThis manual contains specifications, installation instructions, and operating instructions for
Comentarios a estos manuales