Vista de pagina 52
RF Switch Error Messages 53
Appendix C
Appendix C
RF Switch Error Messages
The following error messages are unique to the E1470A. See the
appropriate command module or VXI Controller module manual for a more
complete list of possible error messages.
Error
Number
Message Generated Description Commands that
may cause error
-108 “Parameter not allowed” A parameter was specified that is not valid for the
command. More than 80 channels specified in
command.
-222 “Data out of Range” Invalid numerical state parameter.
-240 “Hardware Error” More than 60 msec was required for the relays to
settle.
2001 “Invalid channel number” Invalid channel number in <source> parameter
2022 “Invalid relay number” Invalid relay number in command
2023 “Invalid common bank
number”
Invalid <comm> parameter in command.
2024 “Invalidsourcebank
number”
Invalid source number in <source> parameter
2025 “InvalidCommon-Source
Combination”
Even though the <comm>and<source>
parameters are valid, the combination is not valid.
The specified source cannot connect to the
specified COM terminal.
Vista de pagina 52