Range
Cable length: 1m
Power (when 50 LOAD is
connected to the test port)
40 dBm to 1 dBm (Frequency 1 GHz)
40 dBm to 0 dBm (Frequency > 1 GHz )
*1
Current (when SHORT is
connected to the test port)
0.0894 mA
rms
to 10 mA
rms
(Frequency 1 GHz)
0.0894 mA
rms
to 8.94 mA
rms
(Frequency > 1 GHz )
Voltage (when OPEN is
connected to the test port)
4.47 mV
rms
to 502 mV
rms
4.47 mV
rms
to 447 mV
rms
Cable length: 2m (when Option 4287A-020 is used)
Power Subtract the following attenuation from the power
(setting value) at 1 m cable length:
Attenuation [dB] = 0.37×
(F: Frequency [GHz])
Resolution
0.1 dB
*2
Uncertainty
Cable length: 1 m
Power (when 50 LOAD is
connected to the test port)
Frequency 1 GHz ± 2 dB (23 ± 5 ºC)
± 4 dB (5 ºC to 40 ºC)
Frequency > 1 GHz ± 3 dB (23 ± 5 ºC)
± 5 dB (5 ºC to 40 ºC)
Cable length: 2 m (when Option 4287A-020 is used)
Power Add 1 dB to the uncertainty at 1 m cable length.
Output impedance 50 (nominal)
F
*1
(Frequency 1 GHz)
(Frequency > 1 GHz )
*1
4
Basic Measurement Characteristics
Oscillator level
Output Impedance
*1. It is possible to set more than 0 dBm (447 mV, 8.94 mA) oscillator level at
frequency > 1 GHz. However, the characteristics at this setting are not guaranteed.
*2. When the unit is set at mV or mA, the entered value is rounded to 0.1 dBm resolution.