7-10 Agilent E5250A Users Guide, Edition 11
Command Reference
*RST
*RST
This command performs an instrument reset. Status after *RST is shown below:
Channel Configuration: Normal Config Mode
Connection Rule: Free
Connection Sequence: Break Before Make
Bias Mode: Off
Bias Port: E5252A: 10
E5255A: None
Bias Enabled Channels: All channels (output ports) are bias enabled
Couple Port Mode: Off
Couple Port: No Couple Port
Card Channel Status: All relays open
Self-Test Result: No Change
Syntax *RST
Example OUTPUT @Hp5250;"*RST"
NOTE This command does not change the self-test result.
For the power-on settings, see :SYSTEM:CPON.