1 Characteristics and Specifications
8 U2761A Service Guide
1 Add 1/10
th
of output amplitude and offset specification per °C for
operation outside the range of 18 °C to 28 °C.
2 DC offset set to 0 V.
3 Spurious output at low amplitude is –70 dBm, typical.
4 Add 1 ppm/°C (average) for operation outside the range of 18 °C to
28 °C.
SWEEP CHARACTERISTICS
Waveforms Sine, Square, Ramp, Arbitrary
Type Linear or Logarithmic
Direction Up or Down
Sweep time 1 ms to 500 s
Trigger Single, External, or Internal
TRIGGER CHARACTERISTICS
Trigger input
Input level TTL compatible
Slope Rising or Falling, Selectable
Pulse width > 100 ns
Input impedance > 10 kΩ, DC coupled
Latency < 500 ns
Jitter (RMS) 6 ns (3.5 ns for Pulse)
Trigger output
Input level TTL compatible into 1 kΩ
Pulse width > 400 ns
Output impedance 50 Ω, typical
Fanout 4 TTL
Rise time 20 ns