Calibration 3
U2761A Service Guide 25
Table 3 - 7 Configuration for AC amplitude flatness (–20 dB range)
verification
U2761A Measurement
Function Amplitude Frequency Nominal Error
Sine 35 mVrms 100.000 kHz 0.000 dB ±0.2 dB
Sine 35 mVrms 160.000 kHz 0.000 dB ±0.35 dB
Sine 35 mVrms 250.000 kHz 0.000 dB ±0.35 dB
Sine 35 mVrms 400.000 kHz 0.000 dB ±0.35 dB
Sine 35 mVrms 630.000 kHz 0.000 dB ±0.35 dB
Sine 35 mVrms 1.00000 MHz 0.000 dB ±0.35 dB
Sine 35 mVrms 1.60000 MHz 0.000 dB ±0.7 dB
Sine 35 mVrms 2.50000 MHz 0.000 dB ±0.7 dB
Sine 35 mVrms 4.00000 MHz 0.000 dB ±0.7 dB
Sine 35 mVrms 6.30000 MHz 0.000 dB ±0.7 dB
Sine 35 mVrms 10.0000 MHz 0.000 dB ±0.7 dB
Sine 35 mVrms 11.5000 MHz 0.000 dB ±0.7 dB
Sine 35 mVrms 13.2000 MHz 0.000 dB ±0.7 dB
Sine 35 mVrms 15.2000 MHz 0.000 dB ±0.7 dB
Sine 35 mVrms 17.4000 MHz 0.000 dB ±0.7 dB
Sine 35 mVrms 20.0000 MHz 0.000 dB ±0.7 dB