7 -4 Agilent Technologies Z5623A Option H48
Advanced Topics
Using an External Computer to Control the Test Set
Write Commands:
info$=“refl_01” ‘command for reflection to port 1
length%=len(info$) ‘length of command
HpibEoi(hHpib;7,0) 'disable EOI
HpibEol(hHpib;7,””,0) 'disable EOL
HpibOutputs (hHpib;712,info$,length%)) 'send command to test set
HpibEol(hHpib;7,chr$(13)+chr$(10),2) 're-enable EOL and set to
chr$(13)+chr$(10)
HpibEoi(hHpib;7,1,) 're-enable EOI hHpib specifies the handle
returned byHpibOpen.
Read Commands:
info$=“id?” ‘command for test set identification.
length%=len(info$) ‘length of command
max.len% = 10 ‘max length data from id? or swxx? function.
infi$ = space$(max.len%)
HpibEoi(hHpib;7,0) 'disable EOI.
HpibEol(hHpib;7,””,0) 'disable EOL
HpibOutputs (hHpib;712,info$,length%)) 'send command to test set.
HpibEnters(hHpib%,712,infi$,max.len%) ‘get data from test set.
HpibEol(hHpib;7,chr$(13)+chr$(10),2) 're-enable EOL and set to
chr$(13)+chr$(10).
HpibEoi(hHpib;7,1,) 're-enable EOI hHpib specifies the handle
returned byHpibOpen.
NOTE
For more information on the EOI and EOL commands, refer to the
programming library manual supplied with the 82335 interface.
Using HPVEE
If you are using HPVEE, be sure to set the Direct I/O in the Advance
Device Configuration so the Read Terminator and Write EOL Sequence
is set to “\n”. In the I/O Transaction make sure the EOL is ON.