5 -28 Agilent Technologies Z5623A Option H48
Performance Verification
Setting Up Limit Testing
Specifying IF Bandwidth and Averaging
Step 1. Click Sweep on the menu bar. When the Sweep menu appears, scroll
down to the IF Bandwidth... button and click it. See Figure 5-28.
Figure 5-28 The “IF Bandwidth...” Button
Step 2. When the IF Bandwidth window pops up as shown in Figure 5-29,
change the IF Bandwidth to 100Hz. (Type, “100<space>”, without the
quotation marks.) The result is shown in Figure 5-30. Click the OK
button.
Figure 5-29 IF Bandwidth Window with Default Value