Module 7
Measurement Functions
7-39
0.5 M ohm (device only)
1.5 M ohm (with Series Resistor)
SMU Series Resistor
The measurement result without the SMU series resistor shows about 0.5 M ohm I-V characteristics.
By setting the SMU series resistor, the measurement result shows about 1.5 M ohm characteristics.
The IV-res setup uses the analysis function and auto analysis function to draw the regression line and
get the slope of the line. For the definition, see the Function Setup screen and the Auto Analysis
Setup screen.