Agilent B1500A Configuration and Connection Guide, Edition 5 3- 9
Connection Guide for Wafer Prober
SMU/GNDU connection with prober
Figure 3-9 MCSMU Output Connection
The High Force/High Sense/Low Force/Low Sense lines should be connected to the DUT
as shown in
Figure 3-9. Wiring shown in Figure 3-14 Cable and Probing needle can be
used. To simplify the connections, omit the wiring of the Common.
For the non-Kelvin connection, use the High Force and the Low Force only. Omit the
wiring of the High Sense and the Low Sense.
Common
Common
Insulator
Plate
Coaxial cable
From N1255A Output
High Force
High Sense
Low Force
Low Sense
To DUT
High terminal
To DUT
Low terminal
Triaxial connector
Wire
High Force
High Sense
Low Force
Low Sense
From N1255A Output