Vista de pagina 93
4- 6 Agilent B1500A Configuration and Connection Guide, Edition 5
Connection and Ordering Examples
Two terminal device
High resolution IV characterization
Figure 4-3 Precision IV configuration
Table 4-7 Target device and application
Table 4-8 Key specification
Device Application Note
Diode I
F
-V
F
, I
R
-V
R
, Breakdown
Resister R-I, R-V 2-wire or 4-wire
Item Range Note
Maximum I/V range 100 mA / 100 V
Minimum I/V resolution 1 fA / 0.5 V
HRSMU
Triaxial cable
Sense
Force
B1500A
HRSMU
Sense
Force
Interlock cable
Connector plate
Fixture/Prober
DUT
AC
Interlock
Coverage of Configuration Example