Agilent Technologies 24A Guía de usuario Pagina 96

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 453
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 95
1-83
Making Measurements
Using Ripple Limits to Test a Device
Figure 1-65 Connections for an Example Ripple Test Measurement
2. Press and choose the measurement settings. For this example, the
measurement settings are as follows:
or on ET models:
You may also want to select settings for the number of data points, power, averaging,
and IF bandwidth.
3. Substitute a thru for the device and perform a response calibration by pressing:
4. Reconnect your test device.
5. To better view the measurement trace, press . Refer to
Figure 1-66.
Preset
Meas
Trans: FWD S21 (B/R)
TRANSMISSN
Center 1.8 G/n
Span 3.4 G/n
Scale Ref
AUTO SCALE
Cal
CALIBRATE MENU
RESPONSE
THRU
Scale Ref
AUTO SCALE
Vista de pagina 95
1 2 ... 91 92 93 94 95 96 97 98 99 100 101 ... 452 453

Comentarios a estos manuales

Sin comentarios