6-2
Calibrating for Increased Measurement Accuracy
How to Use This Chapter
How to Use This Chapter
This chapter is divided into the following subjects:
Calibration Considerations on page 6-4
Procedures for Error Correcting Your Measurements on page 6-10
frequency response error correction
frequency response and isolation error correction
enhanced frequency response error correction
one-port reflection error correction
full two-port error correction (ES Analyzers Only)
Power Meter Measurement Calibration on page 6-33
Calibrating for Noninsertable Devices on page 6-40
adapter removal calibration (ES Analyzers Only)
matched adapters
modify the cal kit thru definition
Calibrating for Non-Coaxial Devices (ES Analyzers Only) on page 6-51
TRL calibration
LRM calibration