Selectable Self
Tests
Seven self tests may be invoked individually via the Self Test menu. The seven
selectable self tests are:
Data Acquisition
RS-232-C
External Trigger BNCs
Keyboard
RAM
ROM
Disk Drive
Cycle through all tests
The required test is selected by moving the cursor to the test and pressing the front
panel SELECT key. A pop-up menu appears with a description of the test to be
performed. The self test does not begin until the cursor is placed on Execute and
the front panel SELECT key is pressed.
After the test has been completed, either Passed, Failed, or Tested is displayed on
the Self Test menu in front of the test. These tests are used as troubleshooting
aids. Section 6 contains information on the individual tests used for
troubleshooting.
Performance
Test Interval
The calibration cycle for the logic analyzer is two years, therefore, periodic
performance verification of the HP 1650B/51B is recommended at two year
intervals. The instrument’s performance should be verified after it has been
serviced, or if improper operation is suspected. Calibration should be performed
before any performance verification tests. Further checks requiring access to the
interior of the instrument are included in the adjustment section, but are not
required for the performance verification.
Test Record The results of the performance tests may be tabulated on the Test Record
provided at the end of this section. The Test Record lists the performance tests
and provides an area to mark whether the pod passed or failed the test. The
results recorded in the table at incoming inspection may be used for later
comparisons of the tests during periodic maintenance, troubleshooting, and after
repairs or adjustments.
HP 1650B/1651B Performance Tests
Service Manual 3-3