434 Chapter7
Language Reference
CF Center Frequency
CF Center Frequency
Syntax
Figure 7-72 CF Syntax
Description
The CF command sets the center frequency and sets the spectrum
analyzer to center frequency and span mode. The span remains
constant, unless it is limited by the spectrum analyzers frequency
range. The start and stop frequencies change as the center frequency
changes.
Parameters
number real from 0 to 2.9E+9 (8560E/EC)
0 to 6.5E+9 (Agilent 8561E//EC)