464 Chapter7
Language Reference
FA Start Frequency
FA Start Frequency
Syntax
Figure 7-107 FA Syntax
Description
The FA command sets the start frequency and sets the spectrum
analyzer to start-frequency and stop-frequency mode. If the start
frequency exceeds the stop frequency, the stop frequency increases to
equal the start frequency plus the minimum span.The center frequency
and span change with changes in the start frequency.
Parameters
number real from 0 to 2.9E+9 (8560E/EC)
0 to 6.5E+9 (Agilent 8561E/EC)