7 Press the Offset (differentiate and integrate) or Ref Levl (FFT) softkey
and rotate the knob closest to the
Cursors key to set the offset
(from the center graticule) or reference level (top graticule) of
the resulting waveform.
Function waveform (F2) is available for viewing, measurement, or storage.
8
Press the Previous Menu softkey.
For FFT functions, an additional menu is available to set additional
parameters. See "FFT Measurement" later in this chapter for more
information.
Function 2 Operating Hints
Timebase must be set to Main (and input channels 3 and 4 to Off on 4-channel
oscilloscopes) when using function 2.
When differential is the operation selected, the value displayed for units per
division and offset is volts per second (V/s). When integral is the operation
selected, the value displayed for units per division and offset is volt seconds
(Vs).
Offset is the value (in volts per second or volt seconds) assigned to the center
graticule for function 2. Normal screen position is 0 offset, or at the center
graticule (until changed).
See "Making Cursor Measurements", and "Saving and Recalling Traces" in this
chapter for more information.
Operating the Measurement/Storage Module
Function 2
2–7