Setting Thresholds
Without the Measurement/Storage module installed, rise time and fall time
measurements are performed at the 10%/90% threshold levels. The
remaining five time measurements (frequency, period, duty cycle, positive
pulse width, and negative pulse width) are all performed at the 50%
transition point. Refer to the User and Service Guide for your oscilloscope
for more information.
With the Measurement/Storage module installed, the thresholds are user
selectable. Rise time and fall time measurements are performed at 10%/90%,
20%/80%, or at a user defined threshold level. The remaining five time
measurements are performed at the center point of the currently selected
upper and lower threshold values.
If 10%/90% is selected, the center is 50%.
If 20%/80% is selected, the center is 50%.
If voltage is selected, the center is dependent on the current lower and
upper values.
As an example, if the lower value is set to 0 V, and the upper value is set to
50 mV, then the 50% level is 25 mV. 25 mV is the point that frequency,
period, duty cycle, positive pulse width, and negative pulse width will be
measured. The point of measurement is dependent on the amplitude of the
input signal.
Operating the Measurement/Storage Module
Setting Thresholds
2–15