Chapter 6: Replacing Assemblies
To remove and replace the AutoProbe assembly
105
Figure 6-9
Access Hole
CAUTION AVOID DAMAGE TO THE RIBBON CABLE AND FACEPLATE!
Do not pry around the edge of the assembly. Doing so may damage the ribbon cable or faceplate.
d Finish removing the AutoProbe assembly by feeding the mylar ribbon cable through the
front panel slot.
Figure 6-10
Pushing Out the AutoProbe Faceplate
3 To replace the AutoProbe assembly, reverse the above procedure.