Lab Exercise: Quantification
Testing the Calibration Table
98
Testing the Calibration Table
1) Load the medium standard's data file, med.d. Go to the Report menu, then
Specify Report... or select the Specify Report Calculation and Print Style
tool.
2) Make the following selections:
Destination = Printer and Screen
Calculate = ESTD
Based on = Area
Report Style = Detail
Select chromatogram output.
3) OK the report specifications and save this to your method.
4) From the Report menu, select Print Report or select the Identify Peaks,
calculate Results & print Report tool.
5) Study the resulting report and modify retention time windows or qualifiers if
necessary. Remember to resave the method if you are satisfied with
report style.