Performance Verification Tests
30. Spurious Responses: Agilent E4401B and E4411B
Figure 2-45 Third Order Intermodulation Distortion Test Setup
3. Perform step 4 through step 29 using the information and entries
from Table 2-69. Then continue with step 30 through step 38.
4. Press
Blue Key, Special, 0, 0 on the signal generator. Set the frequency
to the F1 value for TOI Test 1 in Table 2-69. Set the amplitude to
4dBm.
5. Press
PRESET on the synthesized sweeper. Set the CW frequency of
the synthesized sweeper to the F2 value for TOI Test 1 in
Table 2-69. Then press the following:
POWER LEVEL, –10 dBm
RF Off
6. Enter the power sensor calibration factor for the signal generator
Table 2-69 Test Equipment Settings for Testing TOI
TOI Test
F1
(MHz)
F2
(MHz)
Low Pass
Filter
(MHz)
1 50.0 50.05 50
Option 1DR 50.0 50.05 50