6- 24 Agilent EasyEXPERT User’s Guide Vol. 2, Edition 1
Remote Control Interface
SYSTem Subsystem
SYSTem Subsystem
Describes SYSTem subsystem command provided by the EasyEXPERT remote
control interface.
:SYSTem:ERRor[:NEXT]?
Reads and removes the top item in the error queue, and returns the top code and
message.
Syntax :SYSTem:ERRor[:NEXT]?
Query response response<newline>
response: code,"message"
code: Error code. Data type is NR1.
message: Error message. Data type is SRD.
Multiple responses are listed in the FIFO (first-in-first-out) order, separated by a
comma.
If the queue is empty, the response is +0,"No error".