Agilent Technologies EasyEXPERT Guía de usuario Pagina 227

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 344
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 226
Agilent EasyEXPERT User’s Guide Vol. 1, Edition 1 3- 29
Application Test Definition
Auto Analysis
Tangent line The First Point area is used to specify the measurement point the tangent line passes
through.
Enter a variable name and condition expression to specify the point. This is the same
method as specifying the Marker point.
Regression line The First Point area and the Second Point area are used to specify the range of the
measurement data used for the regression calculation. The following two ways are
available to specify the point.
X-Y Coordinate
Enter the desired X-Y coordinate values or expressions in the X and Y fields.
Data Condition
Enter a variable name and condition expression to specify the point. This is the
same method as specifying the Marker point.