1- 52 Agilent EasyEXPERT Users Guide Vol. 1, Edition 1
Main GUI
Repeat Measurement Setup
Thermo-trigger This area is used to set the thermo-trigger function using the N1265A ultra high
current expander/fixture.
Enable
Thermo-trigger
Enables/disables the thermo-trigger function.
N1265A
Thermo-trigger
List...
Opens the Thermo-trigger List dialog box used to set the
thermo-trigger function. See
“Thermo-trigger List” on page
1-53.
If the thermo-trigger function is enabled, the repeat measurement execution flow is
changed as shown in
Figure 1-5. The following steps are added.
1. After Count is judged, temperature (Temp) is measured.
2. If the measured temperature is in the range of the temperature condition (Center
and Radius), the action (Action) is executed after the delay time (Delay) elapsed.
Or else, the temperature measurement and this judgement are repeated.
Where, the terms in ( ) are the setup items on the Thermo-trigger List dialog box.
Figure 1-5 Repeat Measurement Execution Flow with Thermo-trigger Function