Agilent Technologies HP 70910A Guía de usuario Pagina 120

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 154
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 119
120 Chapter 4
Performance Tests
3. Multiple Response Rejection Test
Performance Tests
11.Set the spectrum analyzer to the appropriate millimeter band by using the following
key sequence:
for an 11974A:
SHIFT
for an 11974Q: SHIFT
for an 11974U: SHIFT
for an 11974V: SHIFT
12.Set the spectrum analyzer controls as follows:
SHIFT, 5, MHz, 0, and dB (sets external mixer conversion loss to 0 dB)
FREQUENCY SPAN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50 kHz
ENTER dB/DIV . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 dB
13.Set the spectrum analyzer reference level by pressing
REFERENCE LEVEL and entering
the corresponding value for the band listed below:
for an 11974A:
40 dBm
for an 11974Q:
40 dBm
for an 11974U:
40 dBm
for an 11974V:
55 dBm
14.Set the spectrum analyzer center frequency by pressing
CENTER FREQUENCY and
entering the corresponding value for the band listed below:
for an 11974A: 26.5 GHz
for an 11974Q: 33 GHz
for an 11974U: 40 GHz
for an 11974V: 50 GHz
15.On the spectrum analyzer, press the following keys:
MARKER
NORMAL
PEAK SEARCH
PRESEL PEAK
16.When the preselector-peak routine is complete, press PEAK SEARCH.
17.Record the marker frequency in Table 4-9 on page 144 as Signal Frequency #1.
18.On the spectrum analyzer, press MARKER
.
Vista de pagina 119
1 2 ... 115 116 117 118 119 120 121 122 123 124 125 ... 153 154

Comentarios a estos manuales

Sin comentarios