74 Chapter 3
Operation
Using 8563A, 8561B, 8560A, or 8560 E/EC-Series Spectrum Analyzers
Operation
12.Change the spectrum analyzer center frequency to the value indicated below, then save
this setting in an unused register:
for an 11974A: 40 GHz
for an 11974Q: 50 GHz
for an 11974U: 60 GHz
for an 11974V: 75 GHz
13.On the rear-panel of the 11974, adjust the corresponding potentiometer for the mixer
listed below until one or both of the green LEDs are lit.
for an 11974A: 40 GHz SET
for an 11974Q: 50 GHz SET
for an 11974U: 60 GHz SET
for an 11974V: 75 GHz SET
14.Repeat step 10 through step 13 until green lights are lit at both frequencies without any
additional adjustments.
Operation
Band Selection
1. If necessary, configure the analyzer for a preselected external mixer by pressing
CONFIG
and setting
EXT MXR PRE UNPR to PRE.
2. Set the LO tuning sensitivity as follows:
For the 8564E or 8565G, press
AUX CTRL, REAR PANEL, and setting V/GHz .25 .50
to
.50
For all other analyzers, press AUX CTRL, REAR PANEL, and .5 V/GHz (FAV)
3. If necessary, mate the 11974 to the spectrum analyzer being used for preselected
external mixing by using the Frequency Tracking Calibration procedure above.
4. Use the following key sequence to enter the external mixing mode and to select the
desired band of operation.
AUX CTRL
EXTERNAL MIXER
FULL BAND
Use the or key until the desired band is reached.
NOTE The above band-selection key sequence must be used to enter the external
mixing band of operation. Do not enter in the center frequency of the band
directly.