Measurements and Math Functions 6
InfiniiVision Oscilloscopes User’s Guide 211
Overshoot
Overshoot is distortion that follows a major edge transition expressed as a percentage of
Amplitude. The X cursors show which edge is being measured (edge closest to the trigger
reference point).
A
UTOMATIC OVERSHOOT MEASUREMENT
Rising edge overshoot =
local Maximum - D Top
Amplitude
----------------------------------------------------------
X 100
Falling edge overshoot =
Base - D local Minimum
Amplitude
-----------------------------------------------------------
X 100
local Maximum
Overshoot
Overshoot
local Minimum
Top
Base