Agilent Technologies InfiniiVision 5000 Series Guía de usuario Pagina 343

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 428
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 342
Mask Test 11
InfiniiVision Oscilloscopes User’s Guide 343
X
Y1
Y2
These four parameters are defined in the Oscilloscope Setup portion of the mask file.
The Y-values (normally voltage) are normalized in the file using the following equation:
Y
norm
= (Y - Y1)/Y
where Y = Y2 - Y1
To convert the normalized Y-values in the mask file to voltage:
Y = (Y
norm
* Y) + Y1
where Y = Y2 - Y1
The X-values (normally time) are normalized in the file using the following equation:
X
norm
= (X - X1)/X
To convert the normalized X-values to time:
X = (X
norm
* X) + X1
Oscilloscope Setup Information
The keywords “setup” and “end_setup” (appearing alone on a line) define the beginning
and end of the oscilloscope setup region of the mask file. The oscilloscope setup
information contains remote programming language commands that the oscilloscope
executes when the mask file is loaded.
Any legal remote programming command can be entered in this section.
The mask scaling controls how the normalized vectors are interpreted. This in turn
controls how the mask is drawn on the display. The remote programming commands that
control mask scaling are:
:MTES:SCAL:BIND 0
:MTES:SCAL:X1 -400.000E-06
:MTES:SCAL:XDEL +800.000E-06
:MTES:SCAL:Y1 +359.000E-03
:MTES:SCAL:Y2 +2.35900E+00
Vista de pagina 342
1 2 ... 338 339 340 341 342 343 344 345 346 347 348 ... 427 428

Comentarios a estos manuales

Sin comentarios