Using the Oscilloscope
To set the horizontal reference point
3-46
To set the horizontal reference point
Click on one of the arrows at the bottom of the waveform display.
The selected horizontal reference is highlighted.
The vertical arrows at the bottom of the display correspond to the left, center, and
right horizontal reference points. This is the position of the trigger if the horizontal
position is set to zero delay. A non-zero horizontal position value is the time relative
to the trigger event at the selected horizontal reference point.
Figure 3-19
Horizontal Reference Points
Click to set the
horizontal
reference to the left
side of the display
Click to set the
horizontal
reference to the
center of the display
Click to set the
horizontal
reference to the
right side of the