Infiniium DSO-X 90000A Series Oscilloscopes
Performance characteristics
Sampling model (continued)
Segmented memory Captures bursting signals at max sample rate without consuming memory during periods of inactivity
Number of segments (Up to 524,288 with option 026)
Maximum time between triggers is 562,950 seconds
Re-arm time: 4.5µs
Maximum memory depth: Up to 4 Gpts in 1/2 channel mode with option 026
Filters
Sin(x)/x Interpolation On/off selectable FIR digital filter. Digital Signal Processing
adds points between aquired data points to enhance
measurement accuracy and waveform display
Hardware Trigger
Sensitivity Internal low
Internal high
Auxiliary
Edge Trigger Bandwidth
Minimum Pulse Width Trigger
Hardware
Software (InfiniiScan)
Level Range
Internal
Auxillary
± 4 div from center screen or ± 4 Volts, whichvever is
smallest
± 5 V, also limit input signal to ± 5V
Sweep Modes
Display jitter
(displayed trigger jitter)
Trigger sources Channel 1, Channel 2, Channel 3, Channel 4, aux, and line
Trigger Modes
Edge
Edge Transition
Edge then Edge (time)
Edge then Edge (Event)
Glitch
Line
Pulse Width
Triggers on a specified slope (rising, falling or alternating between rising and falling) and voltage level on
any channel or auxiliary trigger
Trigger on rising or falling edges that cross two voltage levels in > or < the amount of time specified. Edge
transition setting from 250 ps.
The trigger is qualified by an edge. After a specified time delay between 10 ns to 10 s, a rising or falling edge
on any one selected input will generate the trigger
The trigger is qualified by an edge. After a specified delay between 1 to 16,000,000 rising or falling edges,
another rising or falling edge on any one selected input will generate the trigger.
Triggers on glitches narrower than the other pulses in your waveform by specifying a width less than your
narrowest pulse and a polarity. Triggers on glitches as narrow as 125 ps. Glitch range settings: < 250 ps to <
10 s.
Triggers on the line voltage powering the oscilloscope
Trigger on a pulse that is wider or narrower than the other pulses in your waveform by specifying a pulse
width and a polarity. Triggers on pulse widths as narrow as 125 ps. Pulse width range settings 250 ps to 10
s. Trigger point can be “end of pulse” or “time out”.
Runt Triggers on a pulse that crosses one threshold but fails to cross a second threshold before crossing the first
again. Can be time qualified with minimum setting of 250 ps.