Infiniium DSO-X 90000A Series Oscilloscopes
Performance characteristics
Maximum measurement update
rate
> 50,000 measurement/sec (one measurement turned on)
> 250,000 measurement/sec/measurement (ten measurements turned on)
Measurement Modes Standard, Measure All Edges mode
Waveform Measurements
Voltage
Time
Mixed
Frequency Domain
Level Qualifi cation
Peak to peak, minimum, maximum, average, RMS, amplitude, base, top, overshoot, preshoot, upper,
middle, lower
Rise time, fall time, period, frequency, positive width, negative width, duty cycle, burst width, Tmin, Tmax,
Tvolt, setup time (requires Option 002 or 004, standard on DSA models), hold time (requires
Option 002 or 004, standard on DSA models), channel-to-channel delta time, channel-to-channel phase
Area, slew rate, crossing point
FFT frequency, FFT magnitude, FFT delta frequency, FFT delta magnitude
Any channels that are not involved in a measurement can be used to level-qualify all timing measurements
Eye-diagram measurements Eye height, eye width, eye jitter, crossing percentage, Q factor, and duty-cycle distortion
Jitter analysis measurements
Clock
Data
Timing
Requires Option 002 (or E2681A) or 004 (or N5400A). Standard on DSA Series.
Statistics Displays the current, mean, minimum, maximum, range (max-min), standard deviation, number of
measurements value for the displayed automatic measurements
Histograms
Source
Orientation
Measurements
Waveform or measurement
Vertical (for timing and jitter measurements) or horizontal (noise and amplitude change) modes,
regions are defi ned using waveform markers
Mean, standard deviation, mean ± 1, 2, and 3 sigma, median, mode, peak-to-peak, min, max, total
hits, peak (area of most hits), X scale hits, and X offset hits
Mask Testing Allows pass/fail testing to user-defi ned or Agilent-supplied waveform templates. Automask lets you create
a mask template from a captured waveform and defi ne a tolerance range in time/voltage or screen divi-
sions. Test modes (run until) include test forever, test to specifi ed time or event limit,
and stop on failure. Executes “multipurpose” user setting on failure. “Unfold real time eye” feature
will allow individual bit errors to be observed by unfolding a real time eye when clock recovery is
on. Communications mask test kit option provides a set of ITU-T G.703, ANSI T1.102, and IEEE 802.3
industry-standard masks for compliance testing.
Waveform Math
Number of Functions
Hardware Accelerated Math
Operations
Four
Differential and Common Mode
Absolute value, add, average, Butterworth9, common mode, differentiate, divide, FFT magnitude, FFT
phase, FIR9, high pass fi lter, integrate, invert, LFE9, low pass fi lter (4th-order Bessel Thompson fi lter),
magnify, max, min, multiply, RT Eye9, smoothing, SqrtSumOfSquare9, square, square root, subtract,
versus, and optional user defi ned function (Option 010)
FFT
Frequency Range
Frequency Resolution
Window Modes
DC to 40 GHz (at 80 GSa/s) or 20 GHz (at 40 GSa/s)
Sample rate/memory depth = resolution
Hanning, fl attop, rectangular