Agilent Technologies n3306a Manual de usuario Pagina 86

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 102
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 85
Performance Test and Calibration - B
89
+
_
N330xA Module
Figure B-3. VOLTAGE CALIBRATION
Agilent 3458 or
34401 DVM
+
_
N330xA Module
Figure B-4. RESISTANCE CALIBRATION
Agilent 3458 or
34401 DVM
+S
-S
Agilent 3458 or
34401 DVM
SHUNT
Power
Source
+
_
Power
Source
+
_
+S
-S
If 2 DVM's are not available - move DVM between
+/- sense and shunt as required by program
Be sure sense switch is in RMT position
Be sure sense switch is in RMT position
_
Oscilloscope
Power
Source
+
_
Figure B-5. FREQUENCY / DUTY CYCLE TEST
Current
Probe
+
Vista de pagina 85