105
Table 26. (Options 220/225, 420/425, 520/525 only) Cycle Time vs IF Bandwidth
Applies to the Preset condition (201 points, correction off) except for the following changes:
• CF = 10 GHz
• Span = 100 MHz
• Display off (add 21 ms for display on)
Description Typical
IF Bandwidth
(Hz)
Cycle Time (ms)
a
Cycle Time (ms)
Option 080 enabled
250,000 8.9 37.9
200,000 9.3 39.3
150,000 9.9 40.1
100,000 10.5 41.8
70,000 11.5 43.6
50,000 12.8 45.4
30,000 15.4 50
20,000 18.3 53.9
15,000 21 57.5
10,000 27 65.8
7000 34 75.4
5000 48.5 93
3000 72.8 124
2000 108.8 169
1500 126.8 187.1
1000 272.5
700 357.7
500 460
300 697.7
200 1003.5
150 1307.8
100 1917.6
30 6173.8
10 18214.8
1 181699.2
a
Cycle time includes sweep and retrace time.
Comentarios a estos manuales