54
Table 17. Test Port Input (Continued)
Description Specification Typical
Options
020, 025,
120, 125
Options
220, 225
Options
420, 425
Options
520, 525
Options
020, 025,
120, 125
Options
220, 225
Options
420, 425
Options
520, 525
Test Port Noise Floor
1
(Continued)
1 KHz IF Bandwidth
300 kHz to 3
MHz
3
<-73 dBm -- -- -- <-83 dBm -- -- --
3 MHz to 10
MHz
4
<-83 dBm -- -- -- <-90 dBm -- -- --
10 MHz to 45
MHz
4
<-92 dBm -- -- -- <-96 dBm <-69 dBm <-60 dBm <-60 dBm
45 MHz to 70
MHz
<-92 dBm <-76 dBm <-70 dBm <-70 dBm <-96 dBm -- -- --
70 MHz to 500
MHz
<-92 dBm <-80 dBm <-70 dBm <-70 dBm <-96 dBm -- -- --
500 MHz to 2
GHz
<-92 dBm <-85 dBm <-90 dBm <-90 dBm <-100
dBm
-- -- --
2 GHz to 4 GHz <-92 dBm <-85 dBm <-90 dBm <-90 dBm <-100
dBm
-- -- --
4 GHz to 8 GHz <-92 dBm <-85 dBm <-90 dBm <-90 dBm <-99 dBm -- -- --
8 GHz to 10.5
GHz
<-92 dBm <-85 dBm <-80 dBm <-80 dBm <-99 dBm -- -- --
10.5 GHz to
13.5 GHz
<-87 dBm <-85 dBm <-80 dBm <-80 dBm <-94 dBm -- -- --
13.5 GHz to 20
GHz
-- <-85 dBm <-80 dBm <-80 dBm -- -- -- --
20 GHz to
31.25 GHz
-- -- <-80 dBm <-80 dBm
--
--
-- --
31.25 GHz to
40 GHz
-- -- <-75 dBm <-75 dBm -- -- -- --
40 GHz to 50
GHz
-- -- -- <-70 dBm -- -- -- --