15
Table 7. Test Port Input (Continued)
Description Specification Typical
Options 140/145/146/240/245/246 Options 140/145/146/240/245/246
Test Port Compression – 0.1 dB
Power Compression Power Compression
300 KHz to 10 MHz -- -- +5 dBm --
10 MHz to 1 GHz -- -- +9 dBm --
1 GHz to 12.5 GHz -- -- +10 dBm --
12.5 GHz to 13.5 GHz -- -- +9 dBm
13.5 GHz to 20 GHz -- -- +9 dBm --
Table 7. Test Port Input (Continued)
Description Specification Typical
Options
140/240
Option
145/245
Options
146/246
Options
140/240
Options
145/245
Options
146/246
Trace Noise Magnitude
Ratioed measurement, nominal power at test port.
100 KHz IF bandwidth
300 KHz to 10 MHz -- -- -- .015 dB
rms
.030 dB
rms
.050 dB
rms
10 MHz to 10.5 GHz .006 dB
rms
.008 dB
rms
.016 dB
rms
.004 dB
rms
.005 dB
rms
.013 dB
rms
10.5 GHz to 13.5
GHz
.010 dB
rms
.014 dB
rms
.038 dB
rms
.007 dB
rms
.009 dB
rms
.026 dB
rms
13.5 GHz to 20 GHz .010 dB
rms
.014 dB
rms
.038 dB
rms
.007 dB
rms
.009 dB
rms
.026 dB
rms
600 KHz IF bandwidth
300 KHz to 10 MHz -- -- -- .015 dB
rms
.030 dB
rms
.050 dB
rms
10 MHz to 10.5 GHz -- -- -- .013 dB
rms
.015 dB
rms
.032 dB
rms
10.5 GHz to 13.5
GHz
-- -- -- .017 dB
rms
.023 dB
rms
.063 dB
rms
13.5 GHz to 20 GHz -- -- -- .017 dB
rms
.023 dB
rms
.063 dB
rms
100 KHz IF bandwidth
Measured at Maximum Specified Power
300 KHz to 10 MHz -- -- -- .005 dB
rms
.010 dB
rms
.012 dB
rms
10 MHz to 2 GHz -- -- -- .001 dB
rms
.003 dB
rms
.004 dB
rms
2 GHz to 10.5 GHz -- -- -- .002 dB
rms
.003 dB
rms
.004 dB
rms
10.5 GHz to 13.5 GHz -- -- -- .006 dB
rms
.009 dB
rms
.023 dB
rms
13.5 GHz to 20 GHz -- -- -- .006 dB
rms
.009 dB
rms
.023 dB
rms
Comentarios a estos manuales