Range = Autorange
Resolution = 5½ digits
Measuring DC and AC Voltage
Key points:
¾ You must insert the voltmeter leads across the two points in a circuit
for which you want to measure the voltage. Use the red and black input
jacks as shown below.
¾ You can use the rear panel red and black input jacks also (be sure to
select front or rear).
¾ If you use front and rear, you can easily and quickly select one of two
voltages.
¾ The DMM will Autorange, unless you override it by selecting a range.
¾ You can choose the number of digits displayed, using the DIGITS buttons.
¾ Be sure to select DC V or AC V, as needed.
¾ AC voltage displayed is the true RMS value of the waveform.
¾ AC voltage is accurate to less than 1% up to 100 kHz.
¾ Input impedance for AC V is 1M in parallel with 100 pF (not including
test leads).
¾ Input resistance for DC V is 10M , unless changed to 10G ! for 100
mV, 1 V, & 10 V DC ranges