Temperature Measurement Configuration Commands
MEASure
:TEMPerature? [{TC | THER | FRTD | DEF}][,{<
type
> | DEF}][, 1, {<
resolution
> | MAX | MIN | DEF}]
CONFigure
:TEMPerature [{TC | THER | FRTD | DEF}][,{<
type
> | DEF}][, 1, {<
resolution
> | MAX | MIN | DEF}]
[SENSe:]
FUNCtion "TEMPerature"
FUNCtion?
[SENSe:]
TEMPerature:TRANsducer:TYPE {TCouple | THERmistor | FRTD}
TEMPerature:TRANsducer:TYPE?
[SENSe:]
TEMPerature:TRANsducer:TCouple:TYPE {B | E | J | K | N | R | S | T}
TEMPerature:TRANsducer:TCouple:TYPE?
TEMPerature:TRANsducer:TCouple:RJUNction {<
reference
> | MIN | MAX}
TEMPerature:TRANsducer:TCouple:RJUNction? [{MIN | MAX}]
TEMPerature:TRANsducer:TCouple:RJUNction:TYPE {FIX | INT | THER}
TEMPerature:TRANsducer:TCouple:RJUNction:TYPE?
[SENSe:]
TEMPerature:TRANsducer:FRTD:TYPE {85 | 91}
TEMPerature:TRANsducer:FRTD:TYPE?
TEMPerature:TRANsducer:FRTD:RESistance[:REFerence] <
value
>
TEMPerature:TRANsducer:FRTD:RESistance[:REFerence]?
[SENSe:]
TEMPerature:NULL[:STATe] {OFF | ON}
TEMPerature:NULL[:STATe]?
TEMPerature:NULL:VALue {<
value
> | MIN | MAX}
TEMPerature:NULL:VALue?
[SENSe:]
TEMPerature:NPLCycles {0.02 | 0.2 | 1 | 2 | 10 | 20 | 100 | 200 | MIN | MAX}
TEMPerature:NPLCycles? {MIN | MAX}
UNIT
:TEMPerature {C | CEL | F | FAR | K}
:TEMPerature?
Chapter 4 Remote Interface Reference