Agilent Technologies 34450A Guía de usuario Pagina 105

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 140
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 104
Measurement Tutorial 3
34450A User’s Guide 83
Considering the internal resistance and external lead
resistance total is 0.0125 ohm. If applying a 1 A DC current,
an error of (0.0125 ohm x 1 A) 0.0125 V or 12.5 mV will
occur. This error will be relative by range with the ADC
dynamic range of 1.2 V.
Figure 3-4 Example of measuring voltage and current in dual measure-
ment
The error in measurement will be significant when applying a higher
current.
Vista de pagina 104
1 2 ... 100 101 102 103 104 105 106 107 108 109 110 ... 139 140

Comentarios a estos manuales

Sin comentarios