Agilent Technologies 34450A Guía de usuario Pagina 113

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 140
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 112
Measurement Tutorial 3
34450A User’s Guide 91
Other Primary Measurement Functions
Frequency measurement errors
The multimeter uses a reciprocal counting technique to
measure frequency. This method generates constant
measurement resolution for any input frequency. All
frequency counters are susceptible to errors when measuring
low- voltage, low- frequency signals. The effects of both
internal noise and external noise pickup are critical when
measuring “slow” signals. The error is inversely proportional
to frequency. Measurement errors also occur if you attempt
to measure the frequency of an input following a DC offset
voltage change. You must allow the multimeter’s input to
fully settle before making frequency measurements.
Vista de pagina 112
1 2 ... 108 109 110 111 112 113 114 115 116 117 118 ... 139 140

Comentarios a estos manuales

Sin comentarios