14
at measurement BW = 3 3
at measurement BW = 2 4
at measurement BW = 1 6
Y
ODC
=
at dc bias range = 1 mA 0 [S]
at dc bias range = 10 mA 1 [µS]
at dc bias range = 100 mA 10 [µS]
NOTE
f : frequency in MHz.
V
mV
: oscillator level in mV.
Impedance Measurement Accuracy at 3.5-mm Port of the Agilent 42941A
Equation 3 shows the impedance measurement accuracy [%] at 3.5-mm port of
the 42941A Impedance Probe.
Equation 3. Impedance Measurement Accuracy [%] at 3.5-mm Port of the Agilent 42941A
Where,
E
P
' = E
PBW
+ E
POSC
+ E
P
[%]
Y
O
' = K
BW
Ky
OSC
(Y
ODC
+ Y
O
) [S]
Z
S
' = K
BW
Kz
OSC
Z
S
[]
Y
o
, E
p
, Z
s
: See Figure 3 on page 17.
E
posc
[%] =
at oscillator level >500 mV
at oscillator level >125 mV, 0
500 mV
at oscillator level 125 mV
Ky
osc
=
at oscillator level 500 mV 1
at oscillator level <500 mV 500
V
mV
E = E
p
' +
(
Z
s
'
+ Y
o
'
.
|Z
x
|
)
100
|Z
x
|
(
V
mV
1
)
500
f
100
(0.05 + E
PBW
)
(
125
1
)
V
mV