Agilent Technologies 4294A Manual de usuario Pagina 8

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 32
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 7
8
Measurement Accuracy
Conditions of Accuracy Specifications
Temperature
Four-terminal pair port of the 23 ±5 °C
4294As front panel Beyond 23 ±5 °C, the measurement accuracy is twice as bad
as described.
7-mm port of the 42942A Within ±5 °C from the adapter setup temperature.
Terminal Adapter Measurement accuracy applies when the adapter setup is per-
formed at 23 ±5 °C. When the adapter setup is performed beyond
23 ±5 °C, the measurement accuracy is twice as bad as described.
3.5-mm port of the 42941A Within ±5 °C from the adapter setup temperature.
Impedance Probe Measurement accuracy applies when the adapter setup is per-
formed at 23 ±5 °C. When the adapter setup is performed beyond
23 ±5 °C, the measurement accuracy is twice as bad as described.
Measurement terminal of the Within ±5 °C from the adapter setup temperature.
16048G or 16048H Measurement accuracy applies when the adapter setup is per-
formed at 23 ±5 °C. When the adapter setup is performed beyond
23 ±5 °C, the measurement accuracy is twice as bad as described.
Measurement bandwidth 5
Measurement Accuracy
|Z|, |Y| accuracy ±E [%] (see Equation 1 on page 10, Equation 2 on page 12,
Equation 3 on page 14)
θ accuracy ±E /100 [rad]
L, C, X, B accuracy
at D
x
0.1 ±E [%]
at D
x
>0.1 ±E 1 + D
x
2
[%]
R accuracy
`b
Vista de pagina 7
1 2 3 4 5 6 7 8 9 10 11 12 13 ... 31 32

Comentarios a estos manuales

Sin comentarios