Agilent Technologies 4294A Manual de servicio Pagina 402

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 518
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 401
402 Chapter 16
GPIB Command Reference
PORTL
PORTL
Syntax PORTL <numeric>[M]
PORTL?
Description Sets the port extension compensation amount in electrical length.
Parameters
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response {numeric}<newline><^END>
Corresponding
key
[CAL] - PORT EXTENSION - VALUE (RENGTH)
PORTZ
Syntax PORTZ <numeric>[S]
PORTZ?
Description Sets the port extension compensation amount in time.
Parameters
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response {numeric}<newline><^END>
Corresponding
key
[CAL] - PORT EXTENSION - VALUE (TIME)
<numeric>
Description Port extension compensation amount
Range -10×2.998E8 to 10×2.998E8
Initial value 0
Unit m (meter)
Resolution 1E-6
<numeric>
Description Port extension compensation amount
Range -10 to 10
Initial value 0
Unit s (second)
Resolution 1E-12
Vista de pagina 401
1 2 ... 397 398 399 400 401 402 403 404 405 406 407 ... 517 518

Comentarios a estos manuales

Sin comentarios