Vista de pagina 172
Using N8480 Series Power Sensors 6
N1913/1914A EPM Series Power Meters User’s Guide 151
N8480 Series Sensors Connection Requirements
Table 6 - 2 N8480 Series connection requirements
Sensor Connection Requirements
N8481A
N8482A
These power sensors connect directly to the POWER REF.
N8485A This power sensor requires an APC 3.5 (f) to 50 Ω (m) N-Type adapter
(08485-60005) to connect to the POWER REF.Remove this adapter before
making measurements.
N8487A
N8488A
This sensor requires an APC 2.4 (f) to 50 Ω (m) N-Type adapter
(08487-60001) to connect to the POWER REF. Remove this adapter before
making measurements.