Agilent Technologies 437B Guía de usuario Pagina 179

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 261
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 178
Using N8480 Series Power Sensors 6
N1913/1914A EPM Series Power Meters User’s Guide 157
11 Press
to complete the entry.
12 Connect the power sensor to the signal to be measured.
13 The corrected measurement result is displayed.
Figure 6-7 Calibration factor displayed
Example
To make a measurement on channel A with a power sensor which has a
reference calibration factor of 99.8% and a calibration factor of 97.6% at
the measurement frequency.
1 Disconnect the power sensor from any signal source.
2 Press ,
and the channel softkey.
3 Use the numeric keypad to enter 99.8 in the Ref Cal Factor pop- up
window.
4 Press
to complete the entry.
5 Press and the channel softkey. The Zeroing pop- up is
displayed.
%
Calibration factor
REF CFs
REF CF
%
Zero
Vista de pagina 178
1 2 ... 174 175 176 177 178 179 180 181 182 183 184 ... 260 261

Comentarios a estos manuales

Sin comentarios